A chromatogram is a graph of detector signal against time. Each peak represents a substance leaving the column, its position tells you when it emerged, and its size tells you how much signal it produced. Almost every number on an HPLC-based certificate comes from reading those two properties.
Our HPLC explainer covers how the instrument works. This post assumes that background and focuses on interpretation: what the peaks mean, how purity is calculated, and where a result can mislead.
What a chromatogram is, in one picture
Picture a nearly flat line running left to right. Time is on the horizontal axis and detector response is on the vertical axis. When a substance reaches the UV detector, the line rises into a peak and then falls back.
A clean run for a well-behaved sample shows one large peak for the target compound, with a few much smaller peaks, or none, around it. The flat stretches between peaks are the baseline.
Three things are worth reading on any chromatogram: where the peaks sit, how large they are, and how steady the baseline is.
Retention time: how a peak is identified against a reference standard
Retention time is the time from injection to the top of a peak. Under fixed conditions, a given compound emerges at a very reproducible time.
That reproducibility is what makes identification possible. A certified reference standard of the target compound is run under the same conditions as the sample. If the main peak in the sample appears at the same retention time as the reference standard, that supports the conclusion that the sample's main component is the compound in question.
Two points keep this honest:
- The comparison only means something if both runs used the same method, column and conditions.
- A matching retention time is evidence, not proof. Different substances can occasionally emerge at similar times, which is why identity from retention time is described as supported rather than proven.
Peak area and area-percent purity
The size of a peak is measured as its area, not its height, because area tracks the total signal that passed the detector.
Area-percent purity is a simple ratio. In a sentence: purity % = target peak area divided by total peak area, multiplied by 100.
For example, if the target peak accounts for 98.5 parts of every 100 parts of total detected peak area, the area-percent purity is 98.5%.
This figure has a specific meaning. It describes the share of the detected signal that belongs to the target peak. It does not describe the share of the powder by weight, and it assumes that every detected substance responds to UV light in a broadly similar way. Where responses differ, area percent is an approximation.
Baseline, noise and integration: where results can go wrong
Integration is the step in which software decides where a peak starts and ends and calculates its area. It is where a good chromatogram can still produce a questionable number.
Baseline. A stable, flat baseline lets the software draw peak boundaries consistently. A drifting baseline makes the start and end of a peak ambiguous.
Noise. Random fluctuation in the signal sets a floor. Very small peaks close to the noise level cannot be measured reliably, and the lab has to decide whether they are real.
Integration settings. Thresholds for what counts as a peak, and rules for drawing the line under a peak, change the areas. Two analysts with different settings can report slightly different purity from the same raw data.
Manual adjustments. Sometimes an analyst must correct an automatic boundary. Good practice is to apply consistent rules, review the result against the raw trace, and keep the data traceable.
When you look at a chromatogram, check whether the peak baselines look sensible and whether tiny features are being counted or ignored. Both choices change the final percentage.
Reading impurity peaks and shoulders
Peaks other than the main one are usually read as other detected substances. Their sizes, relative to the target, are what reduce the purity figure.
A few patterns come up often:
- Small separate peaks before or after the main peak: other detected components, each with its own retention time.
- A shoulder on the side of the main peak: a closely related substance that is only partly separated from the target. Software may fold it into the main peak or split it off, which affects purity.
- A broad hump at the start of the run: material that is not retained by the column, often related to the sample solvent rather than the compound.
- Peaks in a blank run (a run with no sample): signals from the system or solvents, not from the sample.
Telling a real component from an artefact is a matter of comparing the sample with blanks and reference runs. This is routine laboratory work, and it is why a chromatogram should be read as a whole rather than as a single number.
What a chromatogram cannot tell you
A chromatogram is informative, but its limits are real.
- Co-eluting substances. Two different compounds can emerge at the same time and appear as one peak. The chromatogram cannot reveal this by itself.
- Compounds with no UV response. If a substance does not absorb light at the detection wavelength, it is invisible. It will not reduce the purity figure, because it was never counted.
- Identity is supported, not proven. A retention-time match to a reference standard supports identity. It is not the same as a mass confirmation.
- Water, salts and counterions. These are typically not seen as peaks, so purity says nothing about net content by weight.
Understanding these limits is part of reading a chromatogram properly. A high purity percentage is a good result, but it answers one question only.
How to read the chromatogram on your ANALITIKA certificate
Each ANALITIKA certificate includes the chromatogram for the tested sample next to the numerical results, such as chromatographic purity. A practical way to read it:
- Find the largest peak and read its position on the time axis. That is the retention time the identity comparison rests on.
- Look at the baseline. It should be steady, without large drifts before or after the peaks.
- Note any additional peaks and shoulders, and see how their sizes compare with the main peak.
- Compare what you see with the reported chromatographic purity. The picture and the number should tell the same story.
- Confirm that the report number and sample ID on the certificate match the sample you submitted.